462 Error Messages  
3138
'VT1539A SCP channel 1 not responding.' Indicates a 
communications error between the VT1539A SCP and 
a Remote Signal Conditioning Unit connected to the 
SCP’s channel 1 input. Check the Data Interface 
connection as well as the power connection to 
that RSCU.
3140
'VT1529A/B data not received properly, cable 
connected?'
3141
'Gain of 0.0 not allowed for VT1529A/B channel.'
3142
'Custom EU out of date, bridge connection was 
EXC when created, now is BRID.'
3143
'Custom EU out of date, bridge connection was 
DRID when created, now is EXC.'
3144
'Channels 15722 to 15731 illegal to send to CVT.' 
Only 502 CVT elements for 512 channels so highest 
10 channels can’t be stored in CVT, use FIFO instead. 
See “ROUTe:SEQuence:DEFine” on page 323.
3145
'Shunt ON/OFF allows only a single channel per 
VT1529A/B.' The shunt cal resistor in a VT1529A/B 
can only be connected to one channel at a time. 
The <ch_list> in the OUTPut:SHUNT contained 
references to more than one channel on the same 
VT1529A/B.
3146
'Communication error reading VT1529A/B calibration 
constants, defaults used.'
3147
'Checksum error reading VT1529A/B calibration 
constants, defaults used.' Do CAL:REMote? and 