Valco electronic Valve Oven ユーザーズマニュアル

ページ / 62
Chapter 2
Valve Installation
Configuring Valves
32
Configuring Valves
The valves in the Valve Oven are gas-actuated. The valves are controlled by 
solenoid valves that are turned on and off by timed events. Each solenoid valve 
uses one timed event. Valve 1 will use timed event # 8, Valve 2 will use timed 
event # 7, and so on. Once a valve occupies a timed event, this timed event is 
removed from the list of external events that can be added to a run table.
You can configure a valve to be a gas sampling valve or a switching valve. 
Gas Sampling Valves
Gas sampling valves have two positions—Load and Inject. These correspond to 
the OFF and ON positions, respectively. Gas sampling valves are plumbed so that 
counter-clockwise position is Load and the clockwise position is Inject
 
Figure 2-7.  Mechanical Stop Positions for Gas Valves
Switching Valves
Switching valves have two positions—On and Off. Switching valves are plumbed 
so that the counter-clockwise position is Off and the clockwise position is On.
NOTE
Typically, the valve nearest to the rear of the Valve Oven will be configured as valve 1.
1. Press CONFIG on the TRACE GC Ultra keypad.
;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
@@
@@




;;
;;
yy
yy
;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
@@@
@@@




;;;
;;;
yyy
yyy
Mechanical stop
Load (OFF) Position
Inject (ON) Position