Installation Model 
7600 
 
 
Teledyne Analytical Instruments
 
30 
2.5.6 Purging the Analyzer 
In general, purging the analyzer is not required unless one of the 
following cases apply: 
•  A combustible gas component is contained in sample gas. 
•  Corrosive gas is contained in the atmospheric air at the 
installation site. 
•  The same gas as the sample gas component is contained in 
the atmospheric air at the installation site. 
In the above situations, the inside of analyzer should be purged 
with instrument air or N
2
. Use a purge flow rate of about 1 L/min. 
Use a filter to remove dust or mist from the purge gas. 
2.5.7 Sample Gas Pressure 
The sample gas pressure at the outlet should be atmospheric 
pressure.  
2.5.8 Example configuration of gas 
sampling system 
A typical system configuration with five component gas analysis 
for monito
ri
ng combustion exhaust gas 
fr
om boiler, refuse incinerator, 
etc. is shown in Figure 2-8. Contact Teledyne for specific application 
system configuration or further information. 
The system shown is comprised of: 
•  Model 7600—with dual measuring unit option and external 
zirconia oxygen sensor. 
•  Gas Extractor—with stainless steel filter of standard mesh 40 
μm equipped with integral heater. 
•  Mist Filter—to remove condensate, mist and dust from the 
sample before it enters the analyzer. 
•  Safety Drain Trap—with two compartments for positive and 
negative pressure. It monitors and adjusts the sample gas 
pressure. 
•  Gas Aspirator—to aspirate the sample gas. 
•  Electronic Gas Cooler—used to dry the sample gas to a dew 
point of approximately 2°C (35.6°F).