Teledyne Welding Consumables 7600 ユーザーズマニュアル

ページ / 126
Installation Model 
7600 
 
 
Teledyne Analytical Instruments
 
32 
2.6 Electrical Connections 
2.6.1 Power Inlet 
The power inlet for the Model 7600 is located on the rear panel as 
shown in Figure 2-9. The proper 3-prong power cord has been supplied 
with your instrument. Connect the power cable to the power inlet. 
Figure 2-9: Model 7600 Electrical Connections 
 
The Model 7600 requires 100-240 VAC 50/60 Hz power.  
Avoid installing this instrument near an electrical unit that 
generates electrical noise such as a high frequency furnace or electric 
welder. If using the instrument near such a noise-generating unit is 
unavoidable, use a separate power line to avoid noise. 
Mount a noise suppressor such as varister or spark killer as shown 
in Figure 2-10 to the noise generating unit when noise is generated from 
relays or solenoid valves. The suppressor must be mounted near the 
noise generating source, or it will have no effect.