Teledyne Camera Accessories T100 ユーザーズマニュアル

ページ / 422
Specifications, Approvals & Compliance 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
30 
Parameter Description 
Environmental Installation 
category (over-voltage category) II; Pollution degree 2 
Operating Temperature  
5 - 40 
o
C (with EPA Equivalency) 
Humidity Range 
0 - 95% RH, non-condensing 
Dimensions HxWxD 
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) 
Weight 
31 lbs (14 kg); 35.7 lbs (16 kg) with internal pump 
1
 As defined by the USEPA. 
2
 Defined as twice the zero noise level by the USEPA. 
Table 2-2: O
2
 Sensor Option Specifications 
Parameter Description 
Ranges 
0-1% to 0-100% user selectable. Dual ranges and auto-ranging supported. 
Zero Noise
<0.02% O
 
Lower Detectable Limit
2
 <0.04% 
O
2
 
Zero Drift (24 hours)
 3
 
<± 0.02% O
2
 
Zero Drift (7 days) 
<±- 0.05% O
2
 
Span Noise
<
±
 0.05% O
2
 
Span Drift (7 days) 
<± 0.1% O
2
 
Accuracy  
(intrinsic error) <± 0.1% O
2
  
Linearity 
<± 0.1 % O
2
 
 
Temp Coefficient 
<± 0.05% O
2
 /°C,  
Rise and Fall Time 
<60 seconds to 95% 
1
 As defined by the USEPA 
2
 Defined as twice the zero noise level by the USEPA 
3
 
Note: zero drift is typically <± 0.1% O
2
 during the first 24 hrs of operation 
Table 2-3: CO2 Sensor Option Specifications 
Parameter Description 
Ranges 
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported. 
Zero Noise
<0.02% CO
Zero Drift (24 hours) 
<± 0.02% CO
2
 
Zero Drift (7 days) 
<± 0.05% CO
2
 
Span Noise
<± 0.1% CO
2
 
Span Drift (7 days) 
<± 0.1% CO
2
 
Lower Detectable Limit
2
 <0.04% 
CO
2
 
Accuracy 
<± (0.02% CO
+ 2% of reading) 
Linearity 
<± 0.1% CO
2
 
Temperature Coefficient 
<± 0.01% CO
2
 /°C 
Rise and Fall Time 
<60 seconds to 95% 
1
 As defined by the USEPA 
2
 Defined as twice the zero noise level by the USEPA 
 
 
 
06807C DCN6650