30
Parameter Description
Environmental Installation
category (over-voltage category) II; Pollution degree 2
Operating Temperature
5 - 40
o
C (with EPA Equivalency)
Humidity Range
0 - 95% RH, non-condensing
Dimensions HxWxD
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
31 lbs (14 kg); 35.7 lbs (16 kg) with internal pump
1
As defined by the USEPA.
2
Defined as twice the zero noise level by the USEPA.
Table 2-2: O
2
Sensor Option Specifications
Parameter Description
Ranges
0-1% to 0-100% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% O
2
Lower Detectable Limit
2
<0.04%
O
2
Zero Drift (24 hours)
3
<± 0.02% O
2
Zero Drift (7 days)
<±- 0.05% O
2
Span Noise
1
<
±
0.05% O
2
Span Drift (7 days)
<± 0.1% O
2
Accuracy
(intrinsic error) <± 0.1% O
2
Linearity
<± 0.1 % O
2
Temp Coefficient
<± 0.05% O
2
/°C,
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
3
Note: zero drift is typically <± 0.1% O
2
during the first 24 hrs of operation
Table 2-3: CO2 Sensor Option Specifications
Parameter Description
Ranges
Ranges
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% CO
2
Zero Drift (24 hours)
<± 0.02% CO
2
Zero Drift (7 days)
<± 0.05% CO
2
Span Noise
1
<± 0.1% CO
2
Span Drift (7 days)
<± 0.1% CO
2
Lower Detectable Limit
2
<0.04%
CO
2
Accuracy
<± (0.02% CO
2
+ 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
06807C DCN6650