9600 Compressor Description 
1-6 
P/N 8040540
C
HELIX  TECHNOLOGY  CORPORATION
-
TI
CRYOGENICS
Cooling Water
The water used to cool the Compressor must meet the 
specifications shown in Table 1-3 for proper system operation.
Figure 1-3: Water Flow Rate versus Pressure Drop
Table 1-3: Cooling Water Specifications
Parameter
Value
Maximum Inlet Temperature
90°F (32°C)
Minimum Inlet Temperature
50°F (10°C)
Flow Rate
2.75 ±1.25 gpm (10.4 ± 4.7 lpm)
Pressure Drop (inlet-to-outlet)
See Figure 1-3
Maximum Inlet Pressure
100 psi (6.9 bars)
Alkalinity
6.0 - 8.0 pH
Calcium Carbonate
< 75 ppm
NOTE: Water conditioning may be required for applications not meeting these 
requirements.