Olicom 9600 ユーザーズマニュアル

ページ / 59
Maintenance
5-2 
P/N 8040540
C
HELIX  TECHNOLOGY  CORPORATION
-
TI
CRYOGENICS
Adsorber Replacement
Use the following procedure to change the adsorber every three years.
1. Set the System Circuit Breaker, on the rear of the 9600       
Compressor, to the OFF position.
2. Remove the 4 screws which secure the rear panel to the Compres-
sor and remove the rear panel.
NOTE: Use two wrenches in Step 3 to prevent loosening the body of the 
coupling.
3. Using a 1-3/16 in. wrench, and a 1-1/8 in. wrench, as shown in Fig-
ure 5-1, disconnect the two self sealing coupling connectors quickly 
to minimize helium leakage.
Figure 5-1: Disconnecting Self Sealing Couplings
4. Using a 7/16 in. (11mm) wrench, remove the adsorber mounting 
bolt as shown in Figure 5-2.
5. Move the adsorber from under the mounting tabs in the base as 
shown in Figure 5-2 and remove the adsorber from the Compressor.
6. Install the replacement adsorber under the mounting tabs and 
secure it into place with the bolt removed during Step 4.
7. Using two wrenches as shown in Figure 5-1, connect the two self 
sealing couplings quickly to minimize helium leakage.
8. Install the Compressor rear panel.
This 1 3/16 in. wrench 
is used to turn the self 
sealing coupling 
connector.
This 1 1/8 in. wrench 
holds the coupling in a 
stationary position.
Side View
To Tighten Coupling Halves
Hold Here
Turn Here
Hold Here
Turn Here