Appendix C - Troubleshooting Procedures
C-2 
P/N 8040540
C
HELIX  TECHNOLOGY  CORPORATION
-
TI
CRYOGENICS
Table C-1: Compressor Troubleshooting Procedures
Problem
Possible Cause
Corrective Action
1)  System circuit breaker 
(CB1) trips immediately 
to the OFF (0) position 
when switched to the ON 
(1) position.
1)  Incorrect phasing of input 
power.
1)   Check phasing of input 
power. Refer to Phase 
Check
 in Section 3.
2)  System (CB1) and Con-
trol Circuit (CB2) circuit 
breakers remain in the 
ON (1) position when 
switched ON but the 
compressor will not run.
1)   No power coming from 
source.
2)   Insufficient power 
3)   Remote control jumper 
plug not in place. This will 
apply only if remote cir-
cuit is not being used.
4)   Improperly wired external 
remote control circuit. 
NOTE: Only applies if 
remote control feature is 
being used.
1)    Check source fuses, cir-
cuit breakers, and wiring 
associated with the 
power source.  Repair as 
needed.
2)   Verify adequate phase-to-
phase input voltage.  
Refer to Table 1-2.
3)  Check to insure that 
remote jumper plug is 
fully seated. See Figure 
1-4 for location.  Refer to 
Compressor Remote 
Connector
 in Section 1 
for more information.
4)   Verify correct installation 
of remote control feature.  
Refer to Table 1-6.