Ohaus ohaus indicator cd-11 ユーザーズマニュアル

ページ / 120
EN-11
CD-11 Indicator
2.6.4 Setup Menu
The CD-11 Indicator SEtuP Menu 
must be entered the first time
 the Indicator is used to set the scale base
parameters to match the Indicator.  Do not attempt to calibrate the Indicator before initially setting up the
SEtuP Menu.  All other menus should be entered and set up the first time the Indicator is used.
Procedure
With the Indicator ON, press and hold the     G/N/T/
Menu
  button until MEnU is
displayed.  When the G/N/T/
Menu
 button is released, CAL is displayed if the
CAL jumper on the PC board is in place.  When the CAL jumper is removed, the
Indicator will not permit calibration.  This jumper should be in place initially.
Press the      PRINT/
Units
 button, SEtuP is displayed.
Press the G/N/T/
Menu
     button, LFtOFF is displayed.  Legal for trade selections
are:
'ON' - LFT (Legal for Trade) is ON
'OFF' - LFT is OFF.
'CAn' - LFT is set for Canada
Press the PRINT/
Units
 button until desired LFT setting is reached.
Press G/N/T/
Menu
     button, 0 2 is displayed.  This is the Zero setting.  Selections
are:
2%:
zero operating range is -2% to +2%.
18%: zero operating range is -2% to +18%,
100%: zero operating range is -2% to +100%.
NOTE
: : 
: If LFT is ON, only 2% and 18% are available.
Press the PRINT/
Units
 button until desired zero setting is reached.
Press the G/N/T/
Menu
     button, CAL Un kg is displayed. This is the calibration
unit setting.  Selections are:
'lb' - calibration unit is lb
'kg' - calibration unit is kg.
Press the PRINT/
Units
 button until desired calibration unit setting is reached.