Installation
3.5 Cell compartments, Cell adapters, Light Shield and Protective Cover 
3.5.1 Cell compartments and adapters
The DR 2800 has two cell compartments (
). Only one 
cuvette/sample cell type at a time can be used for a measurement.
Cell compartment #1
13-mm and 16-mm round cuvettes/vials
Note: Cell compartment #1 contains a barcode reader for 
cuvettes/vials.
Cell compartment #2
Cell compartment #2 uses adapters to accommodate different 
cuvette/sample cell types.
1-inch square or 50-mm rectangular cuvettes/cells (can be 
inserted directly into the cell compartment without using an 
adapter). 
Adapter A: 10-mm square cuvettes/cells
Adapter B: Pour-Thru cells (refer to the instruction sheet 
supplied with the Pour-Thru cell) and multi-path cuvettes/cells
Note: Pour-Thru Cell must be used with Adapter B, not Adapter C. 
Adapter C: One-inch round cuvettes/cells and AccuVac® 
Ampules
Note: One-inch round cuvettes/cells and AccuVac Ampules must be 
used with Adapter C, not Adapter B.
Figure 3  Cell compartments
1
Cell compartment #1
2
Cell compartment #2