Upgrades, Repairs, General Maintenance
8990mnt.fm
Section 7
Note: To extend the life of the 
tubing, rotate the pump tube 90 
degrees in the pump housing after 
approximately 
2
/
3
 of the life has 
been realized (once the tube life has 
been determined through use).
Experience at a particular site will be your best indicator of tubing life. Visually 
inspect the tubing and rollers on a regular basis after initial installation to get a 
feel for what maintenance your site will require. Be sure to replace the pump 
tubing before it splits to prolong the life of the sampler and keep the work area 
free of contamination from the sample liquid.
The proper length of silicone tubing must be used in the pump body. An 
improper length can reduce the life of the tubing and pump rollers. Refer to 
 for the correct length.
Figure 29
Pump Tube Loading
7.2.2 Replacing Pump Tubing
1.
To replace the pump tube, remove the four screws on the pump cover. 
2.
Remove the front cover of the pump housing. 
3.
Remove the existing tubing. Locate the black dot on the new tubing. The 
end of the tube that extends farthest beyond the black dot attaches to the 
stainless steel tubing connector. 
4.
Install the pump tube in the pump housing until the black dots are visible 
just outside the pump body. 
5.
After inserting the new pump tube, reinstall the front cover and secure it 
with the four screws until finger tight.
7.3 Upgrades, Repairs, General Maintenance
Only a qualified technician should service the sampler. For example, steps 
that require knowledge of CMOS electrostatic discharge precautions and 
advanced electronics training and should only be performed by a qualified 
technician. If you need assistance in performing any of the following service 
steps, contact the manufacturer.
To Intake Tubing Connector
5 3/4 in. to Sample Fitting
11 5/8 in.
(Tubing in Pump)