Hach 900 MAX ユーザーズマニュアル
Page 30
Installing the Distributor (Multiple Bottle Operation)
Installing the Distributor (Multiple Bottle Operation)
8990hrd.fm
Section 2
2.6.1 Distributor Arm Alignment
1.
Program the sampler for 24-bottle operation.
2.
Press
START PROGRAM
to set the distributor shaft to the
Bottle #1 position.
3.
Place the arm on the distributor shaft and align the rib on the inside wall of
the control housing skirt.
the control housing skirt.
4.
Secure the arm to the shaft by tightening the
1
/
8
in. hex-head screw,
located on the distributor arm.
Figure 13
Distributor Tubing in Arm
Figure 14
Distributor Installation
1.
Distributor Shaft
2.
Nozzle End (
1
/
8
in. max)
1.
Refrigerator
2.
Distributor Assembly
3.
Distributor Arm
1
2
1
2
3