Hach 900 MAX ユーザーズマニュアル

ページ / 160
Page 66
Submerged Pressure Sensor
8990sensors.fm
Section 4
4.3.2 Submerged Pressure Sensor Programming
1.
From the Main Menu, select 
OPTIONS>LEVEL SENSOR.
2.
Select Submerged Xducer using the 
CHANGE CHOICE
 key and press 
ACCEPT.
4.3.3 Submerged Pressure Sensor Calibration
In sites with harsh conditions (extremes of level, temperature, harsh 
chemicals, etc.) calibration should be performed more often.
1.
From the Main Menu, select 
OPTIONS > ADVANCED OPTIONS > 
CALIBRATION > SUBMERGED PROBE.
2.
Choose the orientation that the sensor will be mounted in the flow stream, 
horizontal or vertical, using the 
CHANGE CHOICE
 key. Press 
ACCEPT
.
3.
Lift the sensor out of the water and hold it in the air in the same orientation 
that you selected in the previous step (horizontal or vertical) (
). 
Then press 
ACCEPT
 to continue.
Figure 18
Lifting the Sensor Out of the Water
4.
Follow either the vertical or horizontal procedure below.
Table 6 Submerged Level Sensor Base Board Connection (J21)
Pin
Signal Description
Wire Color
A
V  +
Red
B
Out +
Yellow
C
Out -
Green
D
Ground
Black
11:00 AM 21 - APR - 01 
CALIBRATION
ACCEPT
ORIENTATION OF
CHANGE
CHOICE
SUBMERGED PROBE:
CANCEL
HORIZONTAL
SELECT APPROPRIATE UNITS
Horizontal
Vertical