Hach 900 MAX ユーザーズマニュアル
Section 4
Page 67
8990sensors.fm
Submerged Pressure Sensor
Vertical Orientation Only
a.
Place the sensor under at least 16 cm (6 in.) of water in a vertical
orientation. Make sure the sensor is stable and not moving around.
Then press
orientation. Make sure the sensor is stable and not moving around.
Then press
ACCEPT
to continue.
b. Carefully measure the depth (D
1
) from the surface of the water to the
first weld mark that encircles the sensor body just above the breather
vent holes (
vent holes (
). The weld mark indicates the location of the
internal diaphragm.
c.
Enter the depth (D
1
) and press
ACCEPT
when done.
Figure 19
Measuring Submerged Depth, Vertical Orientation
Horizontal Orientation Only
Note: Always check the Level
Adjust when reinstalling the sampler
following a calibration.
Adjust when reinstalling the sampler
following a calibration.
a.
Place the sensor under at least 16 cm (6 in.) of water in a horizontal
orientation. Make sure the sensor is stable and not moving around.
Then press
orientation. Make sure the sensor is stable and not moving around.
Then press
ACCEPT
to continue.
b. Measure the depth from the bottom of the bucket to the surface of the
water (D
1
) and enter the value. Press the
ACCEPT
.
Figure 20
Measuring Submerged Depth, Horizontal Orientation
1.
Gray Band
2.
Breather Vents
3.
Detachable Nose Cone
D
1
1
2
3
D
1