Hach 900 MAX ユーザーズマニュアル

ページ / 160
Page 69
8990options.fm
Optional Device Installation
Section 5
Optional Device Installation
This section describes how to setup a rain gauge to the Sampler as well as 
how to connect, program, calibrate, and maintain the optional water 
quality probes:
5.1 Rain Gauge
An external “tipping bucket” rain gauge (Cat. No. 2149) can be connected to 
the Rain Gauge connector of the sampler (
). The rain gauge 
provides a dry contact closure to the sampler.
As rainfall collects in the 20 cm (8 in.) diameter funnel, it is directed into one 
side of a “tipping bucket” assembly. As each bucket fills, it causes the bucket 
assembly to tip and empty out into the bottom of the rain gauge. Each tip of 
the bucket causes a single contact closure to the rain gauge and sends a 
short 12 V dc pulse into pin C of the Rain Gauge connector. Each pulse (tip) 
represents 0.025 cm (0.01 in.) of rain.
Figure 21
Rain Gauge Tipping Bucket
Rain Gauge (
Dissolved Oxygen Probe 
(
pH Probe
(
Conductivity Probe 
(
ORP Probe 
(
Table 7 Rain Gauge Base Board Connections (J5)
Pin
Signal Description
A
+12 V dc source output
B
not used
C
+12 V dc pulse input
D
not used
E
not used
F
not used