Hach 900 MAX ユーザーズマニュアル

ページ / 160
Section 5
Page 73
8990options.fm
ORP Probe
The ORP probe consists of three wires; a pink, black, and red wire. There is 
no temperature sensor on the ORP probe. 
1.
Attach the clear wire to either screw on the terminal strip labeled GLASS. 
2.
Attach the black wire to the REF screw on the other terminal strip.
3.
Attach the red wire to the GND screw on the terminal strip. 
5.3.2 ORP Probe Programming
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
3.
Highlight ORP using the 
UP
 and 
DOWN 
keys, then press 
SELECT
.
4.
Press 
CHANGE CHOICE
 to cycle between Logged and Not Logged, then 
press 
ACCEPT 
to continue.
5.
Enter a logging interval, then press 
ACCEPT
. Valid logging intervals are 
shown on the status bar along the bottom edge of the display. 
6.
Select another channel to configure or press 
RETURN
 to back up one 
step. Press 
MAIN MENU
 to return to the Main Menu. 
5.3.3 ORP Probe Calibration
5.3.3.1 ORP Preamplifier/Junction Box Calibration
Calibration of the ORP input circuit requires a source of dc voltage between 
500 and 2000 m V dc. The reference voltage must be applied to the ORP 
input terminals on the preamplifier/junction box during calibration. A regulated 
dc power supply or a standard “C” cell battery (1500 mV dc) make excellent 
sources for reference voltage.
1.
From the Main Menu, select 
OPTIONS > ADVANCED OPTIONS > 
CALIBRATION > ORP.
2.
Install the ORP junction box on the sampler with the ORP probe removed. 
3.
Apply a positive reference voltage to the ORP probe terminals in the 
junction box, using either a 1.5 V dc “C” cell battery or a regulated 
power supply. 
Table 9 ORP Connector Pin Assignments (J3)
Pin
Signal Description
Wire Color
A
+5  V  dc
White
B
ground
Blue
C
reference
Yellow
D
pH/ORP
Black
E
5  V  dc
Red
F
RTD
Green