Hach 900 MAX ユーザーズマニュアル

ページ / 160
Page 74
Dissolved Oxygen Probe
8990options.fm
Section 5
4.
Attach the positive battery terminal to the terminal block screw labeled 
“glass” and the negative battery terminal to the terminal block screw 
labeled “ref.”
5.
After making all connections, measure the exact voltage on the “C” cell or 
power supply with a voltmeter. Then press a key to continue. The sampler 
displays the message “Waiting for ORP to Stabilize.” 
6.
Once the reading is sufficiently stable enter a new millivolt level.   The “C” 
cell battery should be approximately 1500 mV (or 1.5 V) when new. Enter 
the exact voltage of the current source in millivolts. 
7.
Press 
ACCEPT
 to store the new calibration values.
8.
Disconnect the “C” cell battery or regulated power supply from the ORP 
input terminals.
9.
Reconnect the ORP Probe leads to the input terminals.
5.4 Dissolved Oxygen Probe
5.4.1 Dissolved Oxygen Probe Connection
This connection is for interfacing the optional D.O. probe to the 
D.O./Conductivity Pre-Amp (Cat. No. 3369).
Strain Relief Recommendation
A strain relief is recommended to protect the cable/probe junction during 
application where the sensor will be thrown or tossed into liquid.
5.4.2 Dissolved Oxygen Probe Programming
Note: The membrane thickness 
must be programmed into the 
instrument. 
The instrument uses 
this information to determine if the 
sensor is generating a reasonable 
current. Failure to program this 
value may result in false error 
conditions
1.
From the Main Menu, select 
OPTIONS>ADVANCED OPTIONS>DATALOG.
2.
Highlight Select Inputs using the 
UP
 and 
DOWN 
keys and press 
SELECT
.
3.
Highlight D.O. using the 
UP
 and 
DOWN 
keys, then press 
SELECT
.
4.
Press 
CHANGE CHOICE
 to cycle between Logged and Not Logged, then 
press 
ACCEPT
.
5.
Enter a logging interval, then press 
ACCEPT
. Valid logging intervals are 
shown on the status bar.
6.
Press 
CHANGE CHOICE 
to select the appropriate units (ppm, ppb, mg/L, 
sat). Press 
ACCEPT
 to continue.
7.
Select another channel to configure, press 
RETURN
 to back up one step 
or press 
MAIN MENU
 to return to the Main Menu.
Table 10 D.O. Connections (J20)
Pin
Signal Description
Wire Color
A
DO - (neg)
Green
B
DO + (pos)
Red
C
Thermister
Black
D
Thermister
Yellow