Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル

ページ / 626
OAUTo
Command
:MARKer:OAUTo {MANual|CHANnel<N>,<type>,<level>,
<slope>,<occurrence>}
The OAUTo command  specifies the automatic placement specification for
the O marker.  The first parameter specifies if automarker placement is to be
in the manual mode or on a specified channel.  If a channel is specified, four
other parameters must be included in the command syntax.  The four
parameters are marker type, level, the slope, and the occurrence count.  
<N>
{1|2}
<type>
{ABSolute | PERCent}
<level>
percentage of waveform voltage level, ranging from 10 to 90 of the Vtop to
Vbase voltage or a voltage level
<slope>
{POSitive | NEGative}
<occurrence>
integer from 1 to 100
Example
OUTPUT XXX;":MARKER:OAUTO CHANNEL1,PERCent,50,POSITIVE,5"
Query
:MARKer:OAUTo?
The OAUTo query returns the current settings.
Returned Format
[:MARKer:OAUTo] (MANual|CHANnel<N>,<type>
<level>,<slope>,<occurrence>}<NL>
Example
OUTPUT XXX;":MARKER:OAUTO?"
MARKer Subsystem
OAUTo
32-9