Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル

ページ / 626
RUNTil (Run Until)
Command
:MARKer:RUNTil
{OFF|LT,<time>|GT,<time>|INRange,<time>,<time>|
OUTRange,<time>,<time>}
The RUNTil command allows you to set a stop condition based on the time
interval between the X marker and the O marker.  In repetitive runs, when
the time specification is met, the oscilloscope stops acquiring data and the
advisory "Stop condition satisfied" is displayed on screen. 
<time>
a real number specifying the time in seconds between the X and O markers
Example
OUTPUT XXX;":MARKER:RUNTIL LT,1MS"
Query
:MARKer:RUNTil?
The RUNTil query will return the current Run Until Time X - O setting.
Returned Format
[:MARKer:RUNTil] {OFF|LT,<time>|GT,<time>|INRange,<time>,
<time>|OUTRange,<time>,<time>}<NL>
Example
OUTPUT XXX;":MARKER:RUNTIL?"
MARKer Subsystem
RUNTil (Run Until)
32-11