Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル

ページ / 626
VOTime?
Query
:MARKer:VOTime? CHANNEL<N>
The VOTime query returns the current voltage level of the selected source at
the O marker.
Returned Format
[:MARKer:VOTime]<level><NL>
<N>
{1|2}
<level>
level in volts where the O marker crosses the waveform
Example
OUTPUT XXX;":MARKER:VOTIME? CHANNEL1"
For compatibility with older systems, the OVOLt query functions the same as the
VOTime query.
VRUNs?
Query
:MARKer:VRUNs?
The VRUNs query returns the number of valid runs and the total number of
runs made.  Valid runs are those where the edge search for both the X and O
markers was successful, resulting in valid marker time measurement.
Returned Format
[:MARKer:VRUNs] <valid_runs>,<total_runs><NL>
<valid_runs>
positive integer
<total_runs>
positive integer
Example
OUTPUT XXX;":MARKER:VRUNS?"
MARKer Subsystem
VOTime?
32-16