Agilent Technologies Printer Accessories 1670G ユーザーズマニュアル
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
XAUTo
Command
:MARKer:XAUTo {MANual|CHANnel<N>,<type>,<level>,
<slope>,<occurrence>}
<slope>,<occurrence>}
The XAUTo command specifies the automatic placement specification for
the X marker. The first parameter specifies if automarker placement is to be
in the Manual mode or on a specified channel. If a channel is specified, four
other parameters must be included in the command syntax. The four
parameters are marker type, level, slope, and the occurrence count.
the X marker. The first parameter specifies if automarker placement is to be
in the Manual mode or on a specified channel. If a channel is specified, four
other parameters must be included in the command syntax. The four
parameters are marker type, level, slope, and the occurrence count.
<N>
{1|2}
<type>
{ABSolute | PERCent}
<level>
percentage of waveform voltage level, ranging from 10 to 90 of the Vtop to
Vbase voltage or a voltage level
Vbase voltage or a voltage level
<slope>
{POSitive | NEGative}
<occurrence>
integer from 1 to 100
Example
OUTPUT XXX;":MARKER:XAUTO CHANNEL1,ABS,4.75,POSITIVE,5"
Query
:MARKer:XAUTo?
The XAUTo query returns the current settings.
Returned Format
[:MARKer:XAUTo] {MANual | CHANnel<N>,<type>,
<level>,<slope>,<occurrence>}<NL>
Example
OUTPUT XXX;":MARKER:XAUTO?"
MARKer Subsystem
XAUTo
XAUTo
32-18