Specifications
Welch Allyn Spot Vital Signs LXi
Nellcor Sensor Accuracy Guide
Accuracy specifications are based on controlled hypoxia studies with healthy, 
non-smoking adult volunteers over the specified saturation SpO
2
 range. Pulse oximeter 
SpO
2
 readings were compared to SaO
2
 values of drawn blood samples measured by 
hemoximetry. All accuracies are expressed as + “X” digits. This variation equals + one 
standard deviation (+ 1 SD), which encompasses 68% of the population.
Nellcor Patents
Covered by one or more of the following US Patents: 4,802,486; 4,869,254; 4,928,692; 
4,934,372; 5,078,136; 5,351,685; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364; 
5,865,736; 6,083,172; 6,708,049; Re. 35,122 and foreign equivalents.
Table 2.  OxiMax Sensor Models, Single Patient Use
Sensor Models
SpO
2
 Range
70% to 100%
MAX-AI
1
 
± 2
MAX-PI*
± 2
MAX-II
± 2
MAX-RI
2
± 3.5
1
 The accuracy specification under motion conditions is ± 3. For a definition of motion, contact Nellcor 
Technical Services or your local Nellcor representative.
2
 The accuracy specification has been determined between saturations of 80% to 100%.
Table 3.  OxiCliq Sensor Models, Single Patient Use
Sensor Models
SpO
2
 Range
70% to 100%
OXICLIQ-PI
± 2.5
Table 4.  Reusable Sensor Models
Sensor Models
SpO
2
 Range
70% to 100%
D-YS (Infant to Adult)
± 3
D-YS and D-YSE
± 3.5
D-YS and D-YSPD
± 3.5
DS-100A
± 3
OXI-A/N (Adult/neonate)
Adult: ± 3
Neonate: ± 4
OXI-P/I (Pediatric/infant)
± 3