Welch Allyn Medical Diagnostic Equipment Crib Toy LXi Manual Do Utilizador

Página de 130
66
Specifications
Welch Allyn Spot Vital Signs LXi
Nellcor Sensor Accuracy Guide
Accuracy specifications are based on controlled hypoxia studies with healthy, 
non-smoking adult volunteers over the specified saturation SpO
2
 range. Pulse oximeter 
SpO
2
 readings were compared to SaO
2
 values of drawn blood samples measured by 
hemoximetry. All accuracies are expressed as + “X” digits. This variation equals + one 
standard deviation (+ 1 SD), which encompasses 68% of the population.
Nellcor Patents
Covered by one or more of the following US Patents: 4,802,486; 4,869,254; 4,928,692; 
4,934,372; 5,078,136; 5,351,685; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364; 
5,865,736; 6,083,172; 6,708,049; Re. 35,122 and foreign equivalents.
Table 2.  OxiMax Sensor Models, Single Patient Use
Sensor Models
SpO
2
 Range
70% to 100%
MAX-AI
1
 
± 2
MAX-PI*
± 2
MAX-II
± 2
MAX-RI
2
± 3.5
1
 The accuracy specification under motion conditions is ± 3. For a definition of motion, contact Nellcor 
Technical Services or your local Nellcor representative.
2
 The accuracy specification has been determined between saturations of 80% to 100%.
Table 3.  OxiCliq Sensor Models, Single Patient Use
Sensor Models
SpO
2
 Range
70% to 100%
OXICLIQ-PI
± 2.5
Table 4.  Reusable Sensor Models
Sensor Models
SpO
2
 Range
70% to 100%
D-YS (Infant to Adult)
± 3
D-YS and D-YSE
± 3.5
D-YS and D-YSPD
± 3.5
DS-100A
± 3
OXI-A/N (Adult/neonate)
Adult: ± 3
Neonate: ± 4
OXI-P/I (Pediatric/infant)
± 3