Precision Trading 60 R3.0 ユーザーズマニュアル

ページ / 62
Operator Manual, Supercharger 60 
R3.0 
 
Page 29 of 62 
6.3.4.  RUN 
6.3.4.1.  Start the Charger-Analyzer in the Single Rate mode. The charger will 
start to reduce the charging current when the battery voltage is within 
0.5V of the float level. 
6.3.4.2.  The charge will terminate automatically (CYCLE END) when the 
TOTAL TIME selected is reached. 
6.3.4.3.  The charge will terminate as a FAULT if the MONITOR CIRCUIT 
detects that: 
•  Battery temperature exceeds 45
o
C/113
o
F (BATTERY 
OVERTEMP FAULT). 
•  The actual charge current deviates from the programmed value by 
more than 1A (CURRENT FAULT). 
 
6.4  PEAK VOLTAGE CHARGE: 
6.4.1.  Nominal Battery Voltage 
6.4.1.1.  Enter the nominal battery voltage using the Cell selector (see chart on 
page 19). 
6.4.1.2.  Select transfer from Main to Topping on Peak Voltage or stop on Peak 
Voltage. 
6.4.2.  Charge Currents 
6.4.2.1.  Program the Main and Topping Charge currents as required. 
6.4.3.  Time 
6.4.3.1.  Program the Main and Total Time selectors to allow the battery to 
reach the required charge level (consult battery manufacturers 
specifications).  
6.4.3.2.  Verify that the Timer Speed is consistent with the required elapsed 
time. 
6.4.4.  RUN 
6.4.4.1.  Start the Charger-Analyzer in the Dual Rate mode. If transfer on peak 
is selected, the charger will transfer from Main to Topping charge 
when the battery reaches the peak voltage. If stop on peak is selected, 
the Charger stop upon reaching the peak voltage. 
6.4.4.2.  The charge will terminate automatically (CYCLE END) when the 
TOTAL TIME selected is reached. 
6.4.4.3.  The charge will terminate as a FAULT if the MONITOR CIRCUIT 
detects that: 
•  Battery temperature exceeds 45
o
C/113
o
F (BATTERY 
OVERTEMP FAULT). 
•  The actual charge current deviates from the programmed value by 
more than 1A (CURRENT FAULT).