7. INSTALLATION
7.1 BENCH SPACE.
7.1.1. The Charger-Analyzer system occupies 19" x 17" (48.3 cm x 43.2 cm) of table
top space for the charger and 10" x 25" (25 cm X 63.5 cm) for the Temp-Plate.
7.1.2. Allow also at least 6" (15.2 cm) of separation from the wall and adjacent
equipment, in order to maintain proper air flow (VITAL!).
7.1.3. NOTE 1: In non air-conditioned rooms or if other heat dissipating equipment are
nearby, it is recommended that circulating or extracting fans be used to aid in the
heated air removal.
heated air removal.
7.1.4. NOTE 2: Operation in dusty or otherwise "dirty" air environments will severely
reduce the cooling capacity of the fans and lead to premature failure.
7.2 LINE VOLTAGE.
7.2.1.
Connect the unit to a wall receptacle NEMA 6-30R, 230V (re-wire the
transformers input for 240V or 208V operation) with a 30 AMP capacity. This
line must be a dedicated line, originating at the breakers box and must be
protected with a motor load rated circuit breaker. Sharing of the line with other
equipment may result in erratic operation of the Charger-Analyzer and the other
devices connected to the same line.
transformers input for 240V or 208V operation) with a 30 AMP capacity. This
line must be a dedicated line, originating at the breakers box and must be
protected with a motor load rated circuit breaker. Sharing of the line with other
equipment may result in erratic operation of the Charger-Analyzer and the other
devices connected to the same line.
7.2.2. NOTE: Operation of this unit with a "soft" line or with a line voltage 10% above
or below the nominal 208/230/240VAC will result in erratic operation and may
also lead to equipment damage if large voltage surges occur.
also lead to equipment damage if large voltage surges occur.
7.2.3. Power (Mains) Wiring Diagram
Figure 3 - Line Voltage Wiring, 115V