Andersen Windows & Doors AN74IX ユーザーズマニュアル

ページ / 85
11 
OWNER’S MANUAL 
Diagrammatic Abator Instructions:  
Installing Optional New Abator Cartridge, Resetting Counter 
When the Abator cartridge has been used completely 
(250 sterilization cycles), the screen will display  
ABATOR CARTRIDGE EXPIRED and the alarm will 
sound.   
 
Additional cycles may be completed by bypassing the 
screen.  This is not recommended.  The cartridge 
should be changed prior to starting a new cycle.  
ABATOR CARTRIDGE 
EXPIRED 
 
EXIT                 BYPASS 
Once a new cartridge is installed, the 
counter must be reset. 
 
To reset the counter: 
1.  Hold the left button continuously for 
more than 3 seconds 
 
 
 
 
 
 
 
 
 
 
2.  Press ABATOR button 
 
 
 
 
 
3.  Answer  YES to the question: 
     INSTALL NEW ABATOR 
     CARTRIDGE 
 
When 
left button 
pressed and held 
for more than 3 
seconds. 
AN74 I 
ANPROLENE STERILIZER 
 
PARAMETERS          START 
INSTALL NEW 
ABATOR CARTRIDGE 
 
NO                            YES 
PRESS ABATOR TO 
INSTALL CARTRIDGE, 
OTHERWISE PRESS DIAG 
ABATOR                    DIAG 
 
CARTRIDGE COUNTER  
RESET TO 250 
DIAGNOSTICS 
 
ABATOR ATTACHED 
© Andersen Products Ltd