Once the items to be sterilized have been disassembled, cleaned, dried and wrapped, the ster-
ilization cycle can proceed. 
 
3.1  Prior to loading the sterilizer: (1) Turn the power ON by pressing the right side of the 
black power switch located on the back right corner of the machine.  This will cause the initial 
screen to appear.  (2) Press the button to the right of START on the display screen to initiate 
the self test.
 
 
3.2.  Preparation and loading of the sterilization liner bag: 
a.  Remove a sterilization liner bag from the refill kit and place it 
 inside the sterilizer cabinet with the open end facing you. 
b.  Place the wrapped items for sterilization inside of the steriliza-
tion liner bag. 
c.  Unroll the gas release bag containing the gas ampoule. 
d.  Without opening the gas release bag, gently move the ampoule 
to the center of the gas release bag.  
e.  Place the gas release bag on top of the wrapped items so that you 
can easily manipulate it through the wall of the bag after the 
sterilization liner bag has been closed.  Do not break the am-
poule at this time.
 
f.  Place appropriate indicators inside of the sterilization liner bag. 
 (See Section 5 for more information concerning sterility indica-
tors.) 
AN74 I 
ANPROLENE STERILIZER 
 
 
 
      START 
Upon successful completion of the self test, the following LOAD screen will be displayed: 
LOAD STERILIZER BAG 
CLOSE BAG OVER TUBE 
PURGE BAG 
 EXIT                        PURGE 
SECTION 3 
      Detailed Pictorial Instructions for Running the Sterilization Cycle 