3.11.  The sterilization cycle will then begin. Note: After the ampoule is activated and the 
door is closed, if the operator does not select a cycle length within 5 seconds, a continuous 
alarm will sound to remind the operator to select the appropriate cycle time. 
12  HOUR CYCLE 
STERILIZING 
11:24:15  REMAINING 
3.12.  At the end of the 12 (or 24) hour cycle, the cabinet ventilation pump will stop and the 
purge pump will run for two minutes, flushing Ethylene Oxide from the sterilization liner bag.  
Then the purge pump will stop while the cabinet ventilation pump runs for two minutes.  This 
alternating cycling of the purge and cabinet ventilation pumps will continue for a period of 
two hours.  This will be displayed as shown below and will count down to 00:00:00. 
BREAK AMPOULE 
CLOSE & LOCK DOOR 
SELECT CYCLE LENGTH 
24 HOUR               12 HOUR 
3.9.  Select the length of the sterilization cycle.  For most items, the 12 HOUR cycle will be 
adequate.  
3.10.  When sterilizing lengths of tubing greater than 3 feet (91.5 cm) or loads containing gas 
absorbent materials, the extended length 24 hour cycle should be used.  Large volumes of gas 
absorbent materials (rubber and plastic) also require the use of an additional Anprolene 
ampoule and the extended 24 hour cycle.  To operate the sterilizer using the extended 24 hour 
cycle, press the button to the left of 24 HOUR.  The display will then show STERILIZING 
24:00:00 REMAINING. 
24 HOUR CYCLE 
STERILIZING 
24:00:00  REMAINING 
3.8.  After the initial purge has been completed (timer to 00:00:00), the display instructs the 
operator to: (1) break the ampoule by manipulating the ampoule gas release bag through the 
wall of the sterilization liner bag, (2) close the door, and (3) lock the door and remove the key.  
INITIAL PURGE 
00:01:30  REMAINING 
 
EXIT  
VENTILATING BAG 
 
01:59:59  REMAINING 