Commands
OA 5000 Series User Manual
3 ---27
NOTE
The *LRN? query always returns a string with command headers,
regardless of the setting of the HEADER command. This is because
the returned string is intended to be able to be sent back to the
OA 5000 as a command string. The VERBOSE command can still
be used normally to specify whether the returned headers should
be abbreviated or full length.
*LRN?
might return the string:
:REFERENCE 0.00;:WAVELENGTH 1300;:ATTENUATION:DB
0.00;:DISPLAY DB;:DISABLE 0;:STORE1 0.00;:STORE2 0.00
*OPC
The *OPC (Operation Complete) command generates the operation com-
plete message in the Standard Event Status Register (SESR) when all
pending operations finish. The *OPC? query places the ASCII character “1”
into the Output Queue when all pending operations are finished. The *OPC?
response is not available to read until all pending operations finish. For a
complete discussion of the use of these registers and the output queue, see
page 3 --- 35.
*WAI, ADJusting?
*OPC
*OPC?
*OPC
?
*PSC
The *PSC (Power-On Status Clear) command sets and queries the power-on
status flag that controls the automatic power-on handling of the DESER,
SRER, and ESER registers. When PSC is TRUE, the DESER register is set to
255 and the SRER and ESER registers are set to 0 at power-on. When PSC
is FALSE, the current values in the DESER, SRER, and ESER registers are
preserved in non-volatile memory when power is shut off and restored at
power-on. For a complete discussion of the use of these registers, see
page 3 --- 35.
Examples:
Related Commands:
Syntax: