Tektronix OA 5000 ユーザーズマニュアル

ページ / 134
Commands
Programming
3 ---28
DESE, *ESE, *RST, *SRE.
*PSC <NRf>
*PSC?
<Space>
?
*PSC
<NRf>
<NRf>
is a value in the range from --- 32767 to 32767.
<NRf>
= 0 sets the power-on status clear flag to FALSE, and disables the
power-on clear and allows the OA 5000 to assert SRQ after power-on.
<NRf>
¸ 0 sets the power-on status clear flag TRUE. Sending *PSC 1
therefore enables the power-on clear and prevents any SRQ assertion after
power-on. Using an out-of-range value causes an execution error.
*PSC 0
sets the power-on status clear flag to FALSE.
*PSC?
might return the value 1, showing that the power-on status clear flag is
set to TRUE.
RECall (No Query Form)
The RECall command sets the attenuation of the instrument to either stored
value 1 or 2.
STORe
RECall {1 | 2}
RECall
1
2
<Space>
Related Commands:
Syntax:
Arguments:
Examples:
Related Commands:
Syntax: