7. Align the holes of the boom assembly mounting bracket with those of the boom 
actuator and insert the bolt. Secure with nut.
NOTE: Be sure that the bolt is completely through the holes of the boom assembly mounting 
bracket and the actuator assembly. The boom assembly will pivot easily if the mounting hardware 
is aligned properly when the boom assembly is secured to the mast.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
NOTE: The bottom of the actuator assembly will already be assembled to the mast mounting bracket.
Installing the Leg Actuator to the Base
NOTE: For this procedure, refer to FIGURE 2.4 on page 20.
1. Slide the leg actuator into the slot in the base of the patient lift. See Detail “B”.
2. Perform the following to secure the leg actuator to the pivot bracket (Detail “A”):
 WARNING
Ensure that there is sufficient room to turn patient lift on its side and that floor area 
is clear of debris. Otherwise, injury to personnel or damage to patient lift may 
occur.
A. Turn the patient lift on its side.
B. Position slot in leg actuator over the pivot bracket. 
C. Install the pin through the leg actuator and pivot bracket and secure with hitch pin.
D. Return the patient lift to the upright position.
3. Perform the following to secure the leg actuator to the mast bracket (Detail “B”):
A. Position leg actuator between mast brackets.
B. Move the legs to align the holes in the leg actuator with the holes in the mast 
bracket. 
C. Install the pin through the holes of the leg actuator and mast bracket and secure 
with hitch pin.
4. Plug the pendant control (not shown) into the bottom of the control box.