ACCESSORIES
43
Installing the Reliant Scale
 WARNING
Patient and sling MUST be removed from the lift during ALL installation proce-
dures.
NOTE: For this procedure, refer to FIGURE 8.2.
1. Position the load cell assembly of the Reliant Scale into the boom mounting bracket. 
Refer to Detail ʺAʺ.
NOTE: Use 1/4‐inch nylon washer. 
2. Secure the Reliant Scale to the boom mounting bracket with a shoulder bolt, two nylon 
washers and a locknut. Securely tighten. Refer to Detail ʺAʺ for washer orientation.
NOTE: Ensure the shaft of the shoulder bolt passes through both sides of the boom mounting 
bracket.
3. Insert swivel bar pin with two nylon washers through the swivel bar. Refer to Detail 
ʺBʺ.
4. Align the mounting holes in the swivel bar pin with the mounting holes in the load 
cell assembly. Refer to Detail ʺBʺ.
5. Secure swivel bar pin to the load cell assembly with the provided mounting screw and 
locknut. Securely tighten. Refer to Detail ʺBʺ.
 WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching 
hardware is tightened securely - otherwise injury or damage may occur.
6. Weigh the patient. Refer to Operating the Scale on page 44 and Weighing the Patient 
FIGURE 8.2   Installing the Reliant Scale
Locknut 
Swivel Bar
Nylon 
Washers 
Swivel Bar 
Pin
Mounting 
Screw
Boom
Reliant 
Scale 
Boom 
Mounting 
Bracket
Reliant 
Scale
Nylon 
Washer
Load Cell 
Assembly 
Shoulder 
Bolt
DETAIL “B” 
DETAIL  “A”
Locknut 
Nylon 
Washer