Philips 862474 사용자 설명서

다운로드
페이지 182
Part 1 Troubleshooting Checklist
Troubleshooting  
6-5
The following two steps are to make sure that the monitor’s AC power supply and battery 
work correctly.
What To Do If The Monitor Cannot Be Switched On, AC Powered:
What To Do If The Monitor Cannot Be Switched On, Battery Powered:
Step
Action
1
Ensure the battery has had ample time to charge (at least 15 minutes). A dead 
battery will prevent the monitor from powering On.
2
Check AC fuses in the AC receptacle.
3
Check keypad connector or replace keypad. See 
 for instructions.
4
If monitor still cannot be switched On, replace the Main PCB module. See 
 for instructions.
Step
Action
1
Plug the monitor into an AC power source.
2
Ensure the battery has had ample time to charge (at least 15 minutes). A dead 
battery will prevent the monitor from powering On.
3
Check AC fuses in the AC receptacle.
4
Check keypad connector or replace keypad. See 
 for instructions.
5
If monitor still cannot be switched On, replace the Main PCB module. See 
 for instructions.