Philips 862474 사용자 설명서
Part 1 Troubleshooting Checklist
Troubleshooting
6-5
The following two steps are to make sure that the monitor’s AC power supply and battery
work correctly.
work correctly.
What To Do If The Monitor Cannot Be Switched On, AC Powered:
What To Do If The Monitor Cannot Be Switched On, Battery Powered:
Step
Action
1
Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.
battery will prevent the monitor from powering On.
2
Check AC fuses in the AC receptacle.
3
Check keypad connector or replace keypad. See
for instructions.
4
If monitor still cannot be switched On, replace the Main PCB module. See
for instructions.
Step
Action
1
Plug the monitor into an AC power source.
2
Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.
battery will prevent the monitor from powering On.
3
Check AC fuses in the AC receptacle.
4
Check keypad connector or replace keypad. See
for instructions.
5
If monitor still cannot be switched On, replace the Main PCB module. See
for instructions.