7 Monitoring Pulse Rate
Switching Pulse On and Off
128
Switching Pulse On and Off
To switch a particular pulse numeric on or off, enter the
Setup Pulse
menu via the measurement
setup menu or wave menu of the pulse source. For example, to switch an SpO
2
pulse numeric on or
off,
1
Enter the
Setup Pulse
menu by selecting the Pulse numeric or by selecting
Pulse
in the
Setup SpO2
menu.
2
In the
Setup Pulse
menu, select
Pulse (Pulse source)
to toggle between
On
and
Off
.
To switch the system pulse on or off, in any
Setup Pulse (Pulse Source)
menu, check
which measurement is currently selected as pulse source. Enter the
Setup Pulse
menu for this
pulse source and then switch off the pulse measurement as described.
Using Pulse Alarms
You can change pulse rate alarm limits in the
Setup Pulse
menu accessed via any Pulse source, or
in the
Setup ECG
menu. Changing the alarm limits for a specific Pulse numeric changes the alarm
limits for all pulse rate alarms and heart rate alarms.
Pulse alarms are only generated when the active alarm source is set to Pulse and a pulse source is set as
system pulse.
system pulse.
Selecting the Active Alarm Source: HR or Pulse?
In most cases the HR and Pulse numerics are identical. In order to avoid simultaneous alarms on HR
and Pulse, the monitor uses either HR or Pulse as its active alarm source. To change the alarm source,
select
and Pulse, the monitor uses either HR or Pulse as its active alarm source. To change the alarm source,
select
Alarm Source
in the
Setup ECG
or
Setup Pulse
menu, then select either
• HR: if you want the HR to be the alarm source for HR/Pulse.
• Pulse: If you select Pulse as the active alarm source, the monitor will prompt you to confirm your
choice. Be aware that if you select Pulse as the alarm source, all arrhythmia and ECG HR alarms are
switched off.
switched off.
• Auto: If the
Alarm Source
is set to
Auto
, the monitor will use the heart rate from the ECG
measurement as the alarm source whenever the ECG measurement is switched on and no
ECG
LEADS OFF
INOP is active.
The monitor will automatically switch to Pulse as the alarm source if:
– a Pulse source is switched on and available,
and
– the heart rate becomes unavailable and the
– a Pulse source is switched on and available,
and
– the heart rate becomes unavailable and the
ECG LEADS OFF
inop is active.
The monitor then uses the pulse rate from the measurement currently selected as system pulse.
While Pulse is the alarm source, all arrhythmia and ECG HR alarms are switched off. If the HR
becomes available again, the monitor automatically uses this as alarm source.
While Pulse is the alarm source, all arrhythmia and ECG HR alarms are switched off. If the HR
becomes available again, the monitor automatically uses this as alarm source.