Using Pulse Alarms
7 Monitoring Pulse Rate
129
WARNING
Selecting Pulse as the active alarm source for HR/Pulse switches off the arrhythmia alarms listed in the 
section “ECG, Arrhythmia, and ST Alarm Overview” on page 104, and the heart rate alarms. This is 
indicated by the message 
All ECG Alarms Off
 (unless this has been configured off for your 
monitor), and the crossed-out alarm symbol beside the ECG heart rate numeric. 
High and low pulse rate and extreme bradycardia and extreme tachycardia alarms from pulse are active. 
Alarm Source Selection Disabled
If Alarm Source Selection is disabled, you cannot change the alarm source. If you try to change the 
source, the monitor displays the message 
To activate enter Config and enable 
AlarmSourceSel
. This setting can only be changed in Configuration Mode.
Changing HR/Pulse Alarm Limits 
As Pulse and HR share the same high and low alarm limits, if you change the alarm limit in the 
Setup Pulse
 menu, the high or low alarm limits for HR in the
 Setup ECG
 menu change 
automatically, and vice versa. The only exceptions are caused by a low limit clamp for each 
measurement: the lowest value for Pulse when derived from SpO
2
 is 30 bpm; for HR 15 bpm, and for 
Pressure 25 bpm. 
Extreme Alarm Limits
The extreme rate alarms, Extreme Tachy and Extreme Brady, generated by the active alarm source, 
either HR or Pulse, are set in Configuration Mode by adding a set value to the high and low alarm 
limits. You need to know what value has been configured for your monitor. Changing the high and 
low alarm limits automatically changes the extreme alarm limits within the allowed range.
To see the values added to the high and low limit alarms to create the extreme rate alarms for your 
monitor, in the 
Setup ECG
 menu, see the menu items 
ExtrTachy
 and 
ExtrBrady
.
QRS Tone
The active alarm source is also used as a source for the QRS tone. You can change the tone volume in 
the
 Setup SpO2
 and 
Setup ECG
 menus and the QRS tone modulation in the 
Setup SpO2
 menu.