VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
Programming the VT1422A for Data Acquisition and Control   129
Setting up the Trigger System
Arm and Trigger
Sources
Figure 4-6 shows the trigger and arm model for the VT1422A. Note that 
when the Trigger Source selected is TIMer (the default), the remaining 
sources become Arm Sources. Using ARM:SOUR allows an event to be 
specified that must occur in order to start the Trigger Timer. The default 
Arm source is IMMediate (always armed).
 Selecting the
 Trigger Source
In order to start an instrument operation cycle, a trigger event must occur. 
The source of this event is selected with the TRIGger:SOURce  <source
command. The following table explains the possible choices for <source>.
BUS
EXTernal
HOLD
IMMediate
TTLTrg<n>
SCP Trg
ARM
/TRIGge
r
S
our
ce
T
ri
gge
r
S
o
u
rc
e
S
el
ec
tor
A
rm
S
o
u
rc
e
S
el
ec
to
r
ARM:SOURce <source>
TRIGger:TIMer <interval>
Trigger
Timer
TIMer
Trigger
Enable
TRIGger:SOURce <source>
Internal
Trigger Signal
Trigger
Counter
TRIGger:COUNt <count>
Figure 4-6. Logical Arm and Trigger Model
Parameter Value
 Source of Trigger (after INITiate:
 command)
BUS
 TRIGger[:IMMediate], *TRG, GET (for GPIB)
EXTernal
 "TRG" signal input on terminal module
HOLD
 TRIGger[:IMMediate]
IMMediate
 The trigger signal is always true (scan starts when an 
INITiate:
 command is received).
SCP
 SCP Trigger Bus (future SCP Breadboard)
TIMer
 The internal trigger interval timer (must set Arm source)
TTLTrg<n>
 The VXIbus TTLTRG lines (n=0 through 7)