VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
130 Programming the VT1422A for Data Acquisition and Control  
 NOTES
1. When TRIGger:SOURce is not TIMer, ARM:SOURce must be set to 
IMMediate (the *RST condition). If not, the INIT command will 
generate an error -221,"Settings conflict."
2. When TRIGger:SOURce is TIMer, the trigger timer interval 
(TRIG:TIM <interval>) must allow enough time to scan all channels, 
execute all algorithms and update all outputs or a +3012, "Trigger 
Too Fast" error will be generated during the trigger cycle. See the 
TRIG:TIM command on page 397 for details.
To set the trigger source to the internal Trigger Timer (the default):
TRIG:SOUR TIMER
now select ARM:SOUR
To set the trigger source to the External Trigger input connection:
TRIG:SOUR EXT
an external trigger signal
To set the trigger source to a VXIbus TTLTRG line:
TRIG:SOUR TTLTRG1
the TTLTRG1 trigger line
Selecting Trigger Timer
Arm Source
Figure 4-6 shows that when the TRIG:SOUR is TIMer, the other trigger 
sources become Arm sources that control when the timer will start. 
The command to select the arm source is ARM:SOURce <source>.
The <source> parameter choices are explained in the following table.
 NOTE
When TRIGger:SOURce is not TIMer, ARM:SOURce must be set to 
IMMediate (the *RST condition). If not, the INIT command will generate 
an error -221,"Settings conflict."
To set the external trigger signal as the arm source:
ARM:SOUR EXT
trigger input on 
connector module
Parameter Value
 Source of Arm (after INITiate:
 command)
BUS
 ARM[:IMMediate]
EXTernal
 "TRG" signal input on terminal module
HOLD
 ARM[:IMMediate]
IMMediate
 The arm signal is always true (scan starts when an 
INITiate:
 command is received).
SCP
 SCP Trigger Bus (future SCP Breadboard)
TTLTrg<n>
 The VXIbus TTLTRG lines (n=0 through 7)