VT1422A Command Reference   307
MEASure
This command executes a measurement scan without running defined 
algorithms. This is to keep algorithms from placing values in the FIFO buffer.
The measurement sample interval is 392 µs
Filter settings and states are not changed by this command.
Note also that shunt resistor source and state are left as currently programmed.
After completing the measurements, the instrument is re-configured to the same 
settings that existed before the command was executed.
When Accepted: Not while INITiated
Related Commands: SENSe:STRain:EXCitation, MEAS:VOLT:UNST?
*RST Condition: Channel unstrained values are not affected by *RST. 
However, *RST changes the function for all analog input channels to Voltage. 
When a strain channel is changed back to the strain function with a 
SENS:FUNC:STRAIN... command, the unstrained values for these channels 
will still be in effect. Of course, loss of power will cause the unstrained values 
to be lost.
Returned Value: numeric, number of channel values in FIFO. The type is int16.
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage
MEAS:VOLT:UNST? (@10000:10931)
determine unstrained voltage on 128 
strain channels through VT1539As in 
SCP positions 0 and 1
SENS:DATA:FIFO:RESET
As in most cases, since the values have 
been sent to the strain EU conversion, 
the individual unstrained values do not 
need to be seen.