VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
VT1422A Command Reference   317
OUTPut
OUTPut:SHUNt?
OUTPut:SHUNt?  (@<channel>) 
returns the status of the shunt resistance on the 
specified Strain SCP or RSC channel.
Parameters
Comments
The <channel> parameter must specify a single channel only.
If <channel> specifies a non strain SCP or RSC, a 3007 "Invalid signal 
conditioning plug-on" error is generated.
Returned Value: Returns 1 or 0. The data type is uint16
Related Commands: OUTP:SHUNT
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage
OUTPUT:SHUNt?  (@116)
Check status of shunt resistance on 
channel 16
OUTPUT:SHUNt?  (@10124)
Check status of shunt resistance on 
VT1529A/B channel 0124
OUTPut:SHUNt:SOURce
OUTPut:SHUNt:SOURce  <select>,(@<ch_list>) 
selects the source of the bridge 
shunt resistance for a VT1529A/B Remote Strain Bridge Conditioning unit. The 
VT1529A/B has an internal shunt resistor and also supports an external user supplied 
resistor.
Parameters
Comments
If <ch_list> specifies a non VT1529A/B strain SCP, a 3007 "Invalid signal 
conditioning plug-on" error is generated.
Only one channel on each VT1529A/B needs to be specified since a single 
resistor is used for all channels in the module.
When Accepted: Not while INITiated
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
channel list (string)
100 - 15731
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
select
discrete (string)
INTernal | EXTernal
none
ch_list
channel list (string)
10000 - 15731
none