VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
316 VT1422A Command Reference  
OUTPut
OUTPut:POLarity?
OUTPut:POLarity? (@<channel>) 
returns the polarity on the digital output channel 
in <channel>.
Parameters
Comments
The <channel> parameter must specify a single channel.
Returned Value: returns one of NORM or INV. The type is string.
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)
OUTPut:SHUNt
OUTPut:SHUNt  <enable>,(@<ch_list>) 
adds shunt resistance to one leg of bridge 
on Strain Bridge Completion SCPs and the VT1529A/B Remote Strain Bridge unit. 
This can be used for diagnostic purposes and characterization of bridge response.
Parameters
Comments
If <ch_list> specifies a non strain SCP, a 3007 "Invalid signal conditioning 
plug-on" error is generated.
Only one channel on any one VT1529A/B can be specified in <ch_list>. This 
is because a single resistor is used to shunt each of a VT1529A/Bs 32 channels. 
The <ch_list> may specify one channel on each of several VT1529A/Bs.
When Accepted: Not while INITiated
Related Commands: [SENSe:]FUNCtion:STRain
…, [SENSe:]STRain…
*RST Condition: OUTP:SHUNT  0 on all Strain SCP and RSC channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
OUTP:SHUNT  1,(@116:119)
add shunt resistance at channels 16 
through 19
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
string
100 - 163
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
enable
boolean (uint16)
0 | 1 | ON | OFF
none
ch_list
channel list (string)
100 - 15731
none