VT1422A Command Reference   353
[SENSe]
[SENSe:]FUNCtion:TEMPerature:POST
[SENSe:]FUNCtion:TEMPerature:POST TC,<sub_type>,[<range>,]
(@<ch_list
>) 
 links VT1529B channels to a post-processing temperature EU 
conversion based on the sensor specified in <type> and <sub_type>Note: Do not 
include the thermocouple reference temperature or thermistor excitation channels in 
this command (for that, use the SENS:REF:POST <type>,<sub_type>,<range>, 
(@<exc_ch>),(@<thr_ch>) command).
This command is valid only on the VT1529B and is available with VXIplug&play 
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).
Parameters
Comments
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4 V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 generates an error. Specifying 
0 selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range.
The VT1529B has a fixed gain of 32. Keep this in mind when <range> is set.
When Accepted: Not while INITiated
Related Commands: SENS:REF:POST and SENS:REF:TEMP:POST
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set 
to FBEN on all strain channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
Link two channels to the J type thermocouple temperature conversion
SENS:FUNC:TEMP:POST  TC,J,(@10002,10003)
Link channel 1 to measure reference temperature using 5k thermistor on VT1586A; 
excitation voltage is on channel 0
SENS:REF:POST  THER,5000,(@10000),(@10001)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
sub_type
discrete (string)
  E | J | T
none
range
 numeric (float32)
see comments
V dc
ch_list
channel list (string)
10000 - 15731
none