VT1422A Command Reference   355
[SENSe]
If using amplifier SCPs or RSCs, set them first and keep their settings in mind 
when specifying a range setting. For instance, if the expected signal voltage is 
to be approximately 0.1 V dc and the amplifier SCP for that channel has a gain 
of 8, <range> must be set no lower than 1 V dc or an input out-of-range condition 
will exist.
The *CAL? command calibrates channels based on Sense Amplifier SCP setup 
at the time of execution. If SCP settings are changed, those channels are no 
longer calibrated. *CAL? must be executed again.
When Accepted: Not while INITiated
Related Commands: *CAL?, INPUT:GAIN
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set 
to FBEN on all strain channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
FUNC:VOLT  (@140:163)
Channels 40 - 63 measure voltage in 
auto-range (defaulted)
[SENSe:]REFerence
[SENSe:]REFerence <type>,<sub_type>,[<range>,](@<ch_list>) 
links channel 
in <ch_list> to the reference junction temperature EU conversion based on <type> 
and <sub_type>. When scanned, the resultant value is stored in the Reference 
Temperature Register and by default the FIFO and CVT. This is a resistance 
temperature measurement and uses the on-board 122 µA current source.
Note
The reference junction temperature value generated by scanning the reference 
channel is stored in the Reference Temperature Register. This reference temperature 
is used to compensate all subsequent thermocouple measurements until the register 
is overwritten by another reference measurement or by specifying a constant 
reference temperature with the SENSE:REF:TEMP command. If used, the reference 
junction channel must be scanned before any thermocouple channels. Use the 
SENSE:REF:CHANNELS command to place the reference measuring channel into 
the scan list ahead of the thermocouple measuring channels.