VT1422A Command Reference   359
[SENSe]
Parameters
Comments
Use SENS:FUNC:TEMP:POST to configure channels to measure thermocouples. 
Then use SENS:REF:POST to configure one or more channels to measure an 
isothermal reference temperature. Finally, use SENS:REF:CHAN:POST to 
group the reference channel with its thermocouple measurement channels.
If thermocouple measurements are made through more than one isothermal 
reference panel, set up a reference channel for each. Execute the 
SENS:REF:CHAN:POST command for each reference/measurement channel 
group.
Related commands: SENS:FUNC:TEMP:POST, SENS:REF:POST
*RST Condition: No channels are linked to references.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:FUNC:TEMP:POST TC,J,.0625,(@10008:10015)      J type TCs on channels 8 through 15
SENS:REF:POST  THER,5000,4,(@10000),(@10001)       sense the reference temperature on 
      ch 10001; excitation on ch 10000;
      measure on 4 V range
SENS:REF:CHAN (@10001),(@10008:10015)
      link TC channels to thermistor
ROUT:SEQ:DEF (@10008:10015,10001,10000)
      set up scan list
[SENSe:]REFerence:TEMPerature
[SENSe:]REFerence:TEMPerature  <degrees_c
stores a fixed reference junction 
temperature in the Reference Temperature Register. Use when the thermocouple 
reference junction is kept at a controlled temperature.
Note
This reference temperature is used to compensate all subsequent thermocouple 
measurements until the register is overwritten by another SENSE:REF:TEMP value 
or by scanning a channel linked with the SENSE:REFERENCE command. If used, 
SENS:REF:TEMP must be executed before scanning any thermocouple channels.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ref_channel
channel list (string)
10000 - 15731
none
ch_list
channel list (string)
10000 - 15731
none