VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
VT1422A Command Reference   357
[SENSe]
[SENSe:]REFerence:POST
[SENSe:]REFerence:POST THERmistor,<resistance>,[<range>,]
(@<exc_ch>),(@<thr_ch
>) 
 links the thermistor channel in <thr_ch> to the 
excitation voltage channel in <exc_ch> and to the reference junction temperature EU 
conversion. The excitation channel is automatically set to SENS:FUNC:HVOL and 
the thermistor channel is set to SENS:FUNC:VOLT. The excitation channel must be 
on the same VT1529B as the thermistor channel.
The reference junction temperature EU conversion assumes the divider circuit shown 
in Figure 5-5 on page 173. If a total resistance value other than 400 k
Ω is used, the 
sum of the resistor values can be specified with the SENS:REF:THER:RES command.
Both the thermistor channel and the excitation channel should appear in the scan list 
set by the ROUT:SEQ:DEF command. If the excitation channel is also used for strain 
measurements, it only needs to appear once in the scan list.
This command is valid only on the VT1529B and is available with VXIplug&play 
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).
Parameters
Comments
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4 V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 causes an error. Specifying 0 
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range.
The VT1529B has a fixed gain of 32. Keep this in mind when <range> is set.
Related Commands: SENS:FUNC:TEMP:POST, 
SENS:REF:THER:RES:POST
*RST Condition: No reference assigned
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:REF:POST  THER,5000,4,(@10000),(@10001) sense the reference temperature on ch 
10001; excitation is on ch 10000; 
measure on 4 V range
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
resistance
numeric (float32)
5000 (value is ignored)
ohm
range
 numeric (float32)
see comments
V dc
exc_ch
channel
10000 - 15731
(one channel only)
thr_list
channel list (string)
10000 - 15731
none