VT1422A Command Reference   367
[SENSe]
[SENSe:]STRain:GFACtor
[SENSe:]STRain:GFACtor <gage_factor>,(@<ch_list>) 
specifies the gage factor 
to be used to convert strain bridge readings for the channels specified by <ch_list>.
Parameters
Comments
The <ch_list> parameter must specify the SCP or RSC channel used to sense 
the bridge voltage, not the channel position on a Bridge Completion SCP.
Related Commands: SENSE:STRAIN:GFAC?, SENSE:FUNC:STRAIN
*RST Condition: Gage factor is 2
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:STRAIN:GFAC 3,(@100:107)
set gage factor for channels 0 through 7
SENS:STRAIN:GFAC 2.2,(@10000:10931)
set gage factor for VT1529A/B channels 
0000 through 0931 (128 channels)
[SENSe:]STRain:GFACtor?
[SENSe:]STRain:GFACtor? (@<channel>) 
returns the gage factor currently set for 
the sense channel specified by <channel>.
Parameters
Comments
Returned Value: Numeric value of gage factor. The data type is float32.
The <channel> parameter must specify a single channel only.
Related Commands: STRAIN:GFACTOR
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
STRAIN:GFAC?  (@107)
query gage factor for channel 7
enter statement here
returns the gage factor set by STR:GFAC
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
gage_factor
numeric (float32)
1 - 5
none
ch_list
channel list (string)
100 - 15731
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
channel list (string)
100 - 15731
none