VT1422A Command Reference   369
[SENSe]
[SENSe:]STRain:UNSTrained
[SENSe:]STRain:UNSTrained <unstrained_v>,(@<ch_list>) 
specifies the 
unstrained voltage value to be used to convert strain bridge readings for the channels 
specified by <ch_list>. The VT1529A/B can use the MEAS:VOLT:UNSTrained 
command which automatically measures the unstrained bridge values and sends each 
value to the channels’ EU conversion. This command does not control the output 
voltage of any source.
Parameters
Comments
Use a voltage measurement of the unstrained strain bridge sense channel to 
determine the correct value for <unstrained_v>.
The <ch_list> parameter must specify the channel used to sense the bridge 
voltage, not the channel position on a Bridge Completion SCP.
Related Commands: SENSE:STRAIN:UNST?, SENSE:FUNC:STRAIN
…, 
MEAS:VOLT:UNSTRained
Power-on Condition: Unstrained voltage is zero
*RST Condition: unaffected by *RST
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
STRAIN:UNST .024,(@100)
set unstrained voltage for channel 0
[SENSe:]STRain:UNSTrained?
[SENSe:]STRain:UNSTrained? (@<channel>) 
returns the unstrained voltage 
value currently used for EU conversion for the sense channel specified by <channel>. 
This command does not make a measurement.
Parameters
Comments
Returned Value: Numeric value of unstrained voltage. The data type is float32.
The <channel> parameter must specify a single channel only.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
unstrained_v
numeric (float32)
-16 through +16
volts
ch_list
channel list (string)
100 - 15731
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
channel list (string)
100 - 15731
none