VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
412 VT1422A Command Reference  
Command Quick Reference
OUTPut (continued)
:TTLTrg<n>
[:STATe] ON | OFF
When module triggered, source a VXIbus trigger on TTLTrg<n>
[:STATe]?
Returns whether the TTL trigger line specified by is enabled
:TYPE PASSive | ACTive,(@<ch_list>)
sets the output drive type for a digital channel
:TYPE? (@<channel>)
Returns the output drive type for <channel>
:VOLTage
:AMPLitude <amplitude>,(@<ch_list>)
Sets the voltage amplitude on Voltage Output and Strain SCPs
:AMPLitude? (@<channel>)
Returns the voltage amplitude setting
ROUTe
:SEQuence
:DEFine (@<ch_list>)
Defines the analog scan list.
:DEFine? AIN | AOUT | DIN | DOUT | DEST
Returns comma separated list of channels in analog I|O, dig I|O ch lists. For 
DEST, returns the data destination for each AIN channel; 0 = none, 1 = CVT, 
2 = FIFO, 3 = CVT&FIFO, -1=set by algorithm (writefifo, writecvt, writeboth)
:POINts? AIN | AOUT | DIN | DOUT
Returns number of channels defined in above lists.
SAMPle
:TIMer <num_samples>,(@<ch_list>)
sets the time interval between channel measurements
:TIMer? (@<channel>)
Returns the time interval between channel measurements
[SENSe:]
CHANnel
:SETTling <settle_time>,(@<ch_list>)
Sets the channel settling time for channels in <ch_list>
:SETTling? (@<channel>)
Returns the channel settling time for <channel>
DATA
:CVTable? (@<ch_list>)
Returns elements of Current Value Table specified by <ch_list>
:RESet
Resets all entries in the Current Value Table to IEEE "Not-a-number"
:FIFO
[:ALL]?
Fetch all readings until instrument returns to trigger idle state
:COUNt?
Returns the number of measurements in the FIFO buffer
:HALF?
Returns 1 if at least 32,768 readings are in FIFO, else returns 0
:HALF?
Fetch 32,768 readings (half the FIFO) when available
:MODE BLOCK | OVERwrite
Set FIFO mode.
:MODE?
Return the currently set FIFO mode
:PART? <n_readings>
Fetch <n_readings> from FIFO reading buffer when available
:RESet
Reset the FIFO counter to 0
FREQuency
:APERture <gate_time>,(@<ch_list>)
Sets the gate time for frequency counting
:APERture? (@<channel>)
Returns the gate time set for frequency counting
FUNCtion
Equate a function and range with groups of channels
:CONDition (@<ch_list>)
Sets function to sense digital state
:CUSTom [<range>,](@<ch_list>)
Links channels to custom EU conversion table loaded by DIAG:CUST:MXB 
or DIAG:CUST:PIEC commands
:HVOLtage [<range>,](@<ch_list>)
Links VT1529B high-level input (DCV measurement) channels to custom EU 
conversion table loaded by DIAG:CUST:MXB or DIAG:CUST:PIEC
:REFerence [<range>,](@<ch_list>)
Links channels to custom reference temperature EU conversion table loaded 
by DIAG:CUST:PIEC commands
:TC <type>,[<range>,](@<ch_list>)
Links channels to custom temperature EU conversion table loaded by 
DIAG:CUST:PIEC and performs ref temp compensation for <type>
:FREQuency (@<ch_list>)
Configure channels to measure frequency
:HVOLtage (@<ch_list>)
Links the specified VT1529B channels to return DCV on high-input pins on 
RJ-45 connector for channel
SCPI Command Quick Reference
Command
Description