VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
466 VT1529A/B Verification & Calibration  
Troubleshooting Failures
A failure in the *CAL? command indicates a problem with the VT1422A or 
with one of the installed SCPs.
A failure in the CAL:REMote? command indicates a problem with the 
VT1529A/B, the VT1539A SCP or the cable between the VT1529A/B and 
the VT1539A.
Test V-3: Sense Out
Description
This test verifies the operation of the excitation sense circuitry of the 
VT1529A/B. It measures the excitation voltage with the dummy load 
attached to each channel. This voltage should be one-half the applied 
excitation voltage due to the voltage divider effect between the 20 k
Ω 
resistor in the dummy load and the 10 k
Ω resistors between the excitation 
bus and the Excitation Sense + and - terminals on channel input in the 
VT1529A/B.
Note that the actual measurements will vary based on the actual resistor 
values in the dummy load, so these values should be considered approximate 
when running this test.
Setup
The VT1529A/B must be connected to the VT1422A through the VT1539A 
SCP. The Dummy Load must be connected to the channel(s) being tested.
The Voltage Source must be connected to the appropriate channel bank 
terminals of the Bridge Excitation input on the VT1529A/B.
The signal path through the VT1529A/B is highlighted in Figure C-2.