VT1529A/B Verification & Calibration   467
Procedure
Set the Voltage Source to +5.000 V. Send the MEAS:VOLT:EXCitation? 
(@<ch_list>) command to the VT1422A. The expected measurement value 
is approximately +2.5 V on each channel.
Troubleshooting Failures
Check that the Dummy Load is wired and working correctly and firmly 
seated in the RJ-45 input connector. If all 8 channels of one block of 
VT1529A/B channels are failing, check that the Voltage Source is 
connected to the correct terminals on the Bridge Excitation input on the 
VT1529A/B.
Test V-4: Bridge
Resistors
Description
This test verifies the operation of the internal bridge completion resistors 
and associated switches in the VT1529A/B.
The operation of the internal bridge resistors can be measured by 
unbalancing the bridge circuit. The bridge output voltage is a function of the 
resistor value as given by the following formula:
V
sense
 = V
excitation
 * (10.0 / (10.0 + 7.5 + R) - 0.5)
Figure C-2. Signal Path for Sense Out Test
Buffer
To VT1422A
(VT1539A SCP)
X2
Buffer
RJ-45
X2
2:1
Mux
Cal-
Cal+
8-pin Telcom
Connector (RJ-45)
Cal-
Cal+
350
120
customer
10K
10K
10K
10K
+
-
Optional
Filter
2,10,100Hz
+Excitation Sense
Instrument.
Amplifier
Dynamic Strain
Output
+ Excitation
-   Excitation
Internal
Rcal
50K
Local Rcal
Enable
Remote Rcal
Enable
Int/Ext Rcal
Select
Excitation Sense
from CHs 1-31
32:1
Mux
CH31
CH00
CH01
32:1
Mux
CH31
CH01
CH00
Bridge Sense
from CHs 1-31
External customer
shunt cal resistor
terminals
-R Cal
+R Cal
Current
Limit
Excitation
to CHs 1-7
enable excitation
CHs 0-7
+Excitation
+Sense
-Sense
-R Cal
+R Cal
+Excitation Sense
-Excitation Sense
-Excitation
Wagner
Voltage enable
off for full bridge
-Excitation Sense
X16
Pin 5
Pin 4
Pin 2
Pin 7
Pin 6
Pin 3
Pin 8
Pin 1
20K